High-Content Imaging
We have updated our filters to Semrock filters. See below for new filter configurations. The ImageXpress 5000A is located in CCSR 0116. Please contact us for key card access to this room.
For high-content automated cellular imaging and analysis we are utilizing the Molecular Devices ImageXpress 5000A with the live cell option.

This instrument contains a fully automated inverted epifluorescence microscope that is capable of scanning a 6-, 24-, 96- or 384-well microplate and recording multiple fluorescent images in a few minutes. The ImageXpress Micro is then able to quantify morphological properties through image analysis software. Our current ImageXpress configuration permits the visualization of both fixed cells and live cells, as the instrument has a temperature-, humidity-, and carbon dioxide-controlled chamber. We plan to add an external automated incubator to permit long-term or large-scale studies of cellular events, and a robotic plate stacker, thereby increasing the throughput of this workstation.
ImageXpress 5000A Brochure
ImageXpress 5000A Schedule. Analysis Computer1, Analysis Computer2
To use the ImageXpress 5000A, please contact David E. Solow-Cordero (x5-6002)
Technical Information
| Objectives |
Excitation Filters |
Emission Filters |
Dichroics |
| 40X 0.6 Plan Fluor ELWD |
427_10nm |
472_30nm |
Triband1 |
| 10X 0.3 Plan Fluor |
504_12nm |
542_27nm |
444/521/608 |
| 20X 0.75 Plan Apo |
589_15nm |
632_22nm |
|
| 20X 0.45 Plan Fluor ELWD |
387_11nm |
440_40nm |
Quadband1 |
| 40X 0.9 S Fluor |
485_20nm |
525_30nm |
410/504/582/669 |
| 4X 0.2 S Fluor |
560_25nm |
607_36nm |
|
| |
650_13nm |
684_24nm |
400DCLP |
| |
D340_X |
D510-80nm |
Q455DCLP |
| |
D360_10X |
ND0.5 |
|
| |
D380_X |
ND2.0 |
|
Filter Sets
| Label Name |
Excitation |
Emission |
Dichroic |
| DAPI |
387_11nm |
440_40nm |
Quadband |
| Cyan GFP |
427_10nm |
472_30nm |
Triband |
| Yellow GFP |
504_12nm |
525_30nm |
Triband |
| Texas Red |
560_25nm |
632_22nm |
Quadband |
| optimized yfp |
504_12nm |
542_27nm |
Triband |
| Fura-2-340 |
D340_X |
D510-80M |
400DCLP |
| Fura-2-360 |
D360_10X |
D510-80M |
400DCLP |
| Fura-2-380 |
D380_X |
D510-80M |
400DCLP |
| GFP |
485_20nm |
525_30nm |
Quadband |
| AlexaF430 |
427_10nm |
542_27nm |
Triband |
| Alexa 594 |
589_15nm |
632_22nm |
Triband |
| Alexa 594Quad |
560_25nm |
607_36nm |
Quadband |
| Alexa 647 |
650_13nm |
684_24nm |
Quadband |
| Dronpa |
387_11nm |
525_30nm |
Q455DCLP |
| FRET |
427_10nm |
525_30nm |
Q455DCLP |
|
|
 |
CFP/YFP/HcRed-3X3M-A Triple-band "Sedat" Filter Set |
|
|
| |
| Optimized for these fluorophores |
Cyan: CFP, AmCyan, SYTOX Blue, BOBO-1, BO-PRO-1
Yellow: YFP, Calcium Green-1, Fluo-3, Rhodamine 123
Red: HcRed, Cy3.5, Texas Red®, Alexa Fluor 594 |
| Complete list of fluorophores for this filter set |
|
|
Actual measured data from typical filters |
|
| |
| This 7-filter "Sedat" set is ideal for high-speed sequential imaging CFP, YFP, and HcRed with high brightness, extremely low crosstalk, and superb signal-to-noise ratio. The set is also ideal for samples triple-labeled with CFP, YFP, and an orange or red dye, like Texas Red. It is comprised of three single-band exciters and three single-band emitters, along with a triple-band beamsplitter, and thus works best when used with both exciter and emitter filter wheels. |
|
| |
| Filter |
Average Transmission |
Center Wavelengths |
Bandwidths * |
| Exciter 1 |
> 90% |
427 nm |
10 nm |
| Exciter 2 |
> 90% |
504 nm |
12 nm |
| Exciter 3 |
> 90% |
589 nm |
15 nm |
| Emitter 1 |
> 90% |
472 nm |
30 nm |
| Emitter 2 |
> 90% |
542 nm |
27 nm |
| Emitter 3 |
> 90% |
632 nm |
22 nm |
| |
Average Reflection |
Reflection Bands |
Transmission Bands |
| Dichroic |
> 95%
> 95%
> 95% |
420 - 430 nm
496 - 510 nm
579 - 596 nm |
451 - 480 nm
530 - 561 nm
618 - 664 nm |
|
|
 |
DA/FI/TR/Cy5-4X4M-B Quad-band "Sedat" Filter Set |
|
|
| |
| Optimized for these fluorophores |
Blue: DAPI, Hoechst, AMCA, Alexa Fluor 350
Green: FITC, GFP, rsGFP, BODIPY, Alexa Fluor 488
Orange: TRITC, Cy3™, Texas Red®, MitoTracker Red,
Alexa Fluor 568 & 594
Red: Cy5™, APC, TOTO-3, TO-PRO-3, Alexa Fluor 647 & 660 |
| Complete list of fluorophores for this filter set |
|
|
Actual measured data from typical filters |
| |
|
|
 |
| |
Exciter 1 |
Dichroic |
Emitter 1 |
| |
|
|
Exciter 2 |
|
Emitter 2 |
| |
|
|
Exciter 3 |
|
Emitter 3 |
| |
|
|
Exciter 4 |
|
Emitter 4 |
| |
|
|
| |
|
| |
| This 9-filter "Sedat" quad-band set is designed for high-speed, sequential imaging of DAPI, FITC, TRITC, and Cy5 (and other key fluorophores – see complete list above). The complete set is comprised of a quad-band beamsplitter with four single-band emitters and four “no burn-out” single-band exciters. The single-band filters are intended to be mounted in filter wheels. |
|
| |
| Filter |
Average Transmission |
Center Wavelengths |
Bandwidths * |
| Exciter 1 |
> 85% |
387 nm |
11 nm |
| Exciter 2 |
> 90% |
485 nm |
20 nm |
| Exciter 3 |
> 90% |
560 nm |
25 nm |
| Exciter 4 |
> 90% |
650 nm |
13 nm |
| Emitter 1 |
> 90% |
440 nm |
40 nm |
| Emitter 2 |
> 90% |
525 nm |
30 nm |
| Emitter 3 |
> 90% |
607 nm |
36 nm |
| Emitter 4 |
> 90% |
684 nm |
24 nm |
| |
Average Reflection |
Reflection Bands |
Transmission Bands |
| Dichroic |
> 95%
> 95%
> 95%
> 95% |
381 - 392 nm
475 - 495 nm
547 - 572 nm
643 - 656 nm
|
420 - 460 nm
510 - 531 nm
589 - 623 nm
677 - 722 nm |
|
| |
|